Report of a case of disorder in patient extubation due to improper fixing of the endotracheal tube

Document Type : Case Report

Authors

1 Assistant Professor, Department of Internal Medicine, Jahrom University of medical sciences, Jahrom, Iran.

2 Associated Professor, Department of Anesthesiology, Jahrom University of medical sciences, Jahrom, Iran.

Abstract

Introduction: Difficult tracheal extubation is defined as unsuccessful or incomplete attempts at targeted removal of
an endotracheal tube (ETT). Difficult extubation is a rare and dangerous complication after surgery or in the intensive care unit. The present study has reported a case of disorder in patient extubation following inappropriate fixation of the endotracheal tube.
Case Personation: The patient is a 42-year-old man who underwent laminectomy surgery due to lumbar disc problems.
There was no specific disease in the previous history of the patient. The patient was intubated with a size 8 spring tracheal tube and the tube cuff was filled with air according to the standard. The surgery was performed in the prone position. After the end of the operation, the patient turned back to the supine position and the process of waking up the patient began. After reversing the effect of relaxants by atropine and neostigmine at the appropriate time, the tube was removed. The cuff of the tube was emptied with a 10 cc syringe, but the tube could not be removed. The anesthesiologist tried to remove the spring tube several times, but it was not possible to extubate the patient. At first, it was assumed
that the tube would break and the metal spring would come out and the spring would get stuck in the trachea. Finally, with double force, this tube was removed while its cuff was full of air, and after investigation, the reason was that the air drain tube was folded by the glue that was used to fix the tube on the patient's face.
Conclusion: Based on the results of the present study, improper fixation in patients undergoing general anesthesia
can increase the risk of difficult extubation by disrupting the process of emptying the tracheal tube cuff.

Keywords

1. Beloncle F, Lorente JA, Esteban A, Brochard L.
Update in acute lung injury and mechanical
ventilation 2013. Am J Respir Crit Care Med. 2014;
189:1187–93.
2. Sugiyama, K. and Yokoyama, K. Displacement of the
Endotracheal Tube Caused by Change of Head
Position in Pediatric Anesthesia: Evaluation by
Fiberoptic Bronchoscopy. Anesth Analg.1996; 82,
251-53.
3. Kaplow R, Bookbinder M. A comparison of four
endotracheal tube holders. Heart Lung. 1994; 23(1):
59-66
4. Mohammed HM, Hassan MS. Endotracheal tube
securements: effectiveness of three techniques among
orally intubated patients. Egyptian Egypt J Chest Dis
Tuberc. 2015; 64(1):183-96.
5. Song YG, Yun EK. Model for unplanned self
extubation of ICU patients using system dynamics
approach. J Korean Acad Nurs. 2015;45(2):280-92.
6. Cavallone LF, Vannucci A. Extubation of the
difficult airway and extubation failure. Anesth Analg.
2013;116(2):368–83.
7. Sorbello M, Frova G. When the end is really the end?
The extubation in the difficult airway patient. Minerva
Anestesiol. 2012;79(2):194-9.
8. Nakagawa H, Komatsu R, Hayashi K. Fiberoptic
evaluation of the difficult extubation. Anesthesiology.
1995;82: 785-86.
9. Nag DS, Samaddar DP. Inappropriate fixation of an
endotracheal tube causing cuff malfunction resulting
in difficult extubation. BJAN. 2016;66(5):536-8.
10. Merlin T, Mileham J. An unusual complication
related to the deflation of two endotracheal tube cuffs.
Vet Anaesth Analg. 2020 May;47(3):414-415.
11. Suman S, Ganjoo P, Tandon MS. Difficult
extubation due to failure of an endotracheal tube cuff
deflation. J Anaesthesiol Clin Pharmacol.
2011;27(1):141-2.
12. Hulme J, Agarwal S. Failure to deflate an
endotracheal cuff. Acta Anaesthesiol Scand.
2008;52:719.
13. Adhikary SD, George SP, Korula G. Failure of
endotracheal cuff deflation. Acta Anaesthesiol Scand.
2005;49:590.
14. Norgate D, Jimenez CP. A rare complication
associated with the endotracheal tube during
extubation in a cat. Vet Anaesth Analg.
2017;44(6):1401-03.15. Panda CK, Karim HMR. Too much of anything is
bad: An unusual case of a stuck endotracheal tube with
deflated cuff. Saudi J Anaesth. 2019;13(2):156-157.
16. Sanchis Mora S, Seymour C. An unusual
complication of endotracheal intubation. Vet Anaesth
Analg. 2011;38(2):158-9.