Comparison of Various Blood Pressure Measurement Techniques with Standard Office Based Blood Pressure Mesurement
Mahsa Ahadi, Alireza Abdi, Armin Attar Dr.
Department of Cardiovascular Medicine, Shiraz University of Medical Sciences, Shiraz, Iran
Abstract:   (546 Views)
Introduction: In order to control Hypertension, it is necessary to understand the methods of blood pressure measurement. Recent guidelines emphasize on employing appropriate blood pressure measurement techniques and white coat hypertension. Definition and methods of blood pressure measurement have changed in the American guideline based on SPRINT trial (Systolic Blood Pressure Intervention Trial), though Europeans still insist on the standard method of office blood pressure measurement. The differences made by SPRINT method are not fully understood yet. This article compares the measurements done by mercury sphygmomanometers and digital devices, and also evaluates the effect of applying the SPRINT method on these readings.
Method: 87 patients were enrolled in this study, For all of whom blood pressure was measured in a quiet room after 5 minutes of resting for 3 times (using an automated device for the first 2 times and a mercury sphygmomanometer for the third time). The mean of them was considered as SPRINT like blood pressure. The results were compared to the measurements done by a cardiologist and a nurse using an automated device at the same visit.
Result: Office based SBPs were about 8.47 mmHg higher than SPRINT like measurements (p<0.001). Diastolic blood pressures were about 7.05 mmHg higher in office (p<0.001).
Conclusion: the methods employed to measure the blood pressure have a pivotal effect on the results. It seems that further studying is needed in this field in Iran.
Keywords: blood pressure, measurement, mercury sphygmomanometer, digital, SPRINT
Full-Text [PDF 609 kb]   (119 Downloads)    
Type of Study: Research | Subject: ect
Received: 2019/09/11 | Accepted: 2020/04/5 | Published: 2020/04/5
Send email to the article author

Add your comments about this article
Your username or Email:

CAPTCHA


XML   Persian Abstract   Print


Download citation:
BibTeX | RIS | EndNote | Medlars | ProCite | Reference Manager | RefWorks
Send citation to:

Ahadi M, Abdi A, Attar A. Comparison of Various Blood Pressure Measurement Techniques with Standard Office Based Blood Pressure Mesurement. jmj. 2020; 18 (1) :1-8
URL: http://jmj.jums.ac.ir/article-1-1392-en.html


Volume 18, Issue 1 (spring 2020) Back to browse issues page
فصلنامه دانشگاه علوم پزشکی جهرم Journal of Jahrom University of Medical Sciences
Persian site map - English site map - Created in 0.06 seconds with 31 queries by YEKTAWEB 4162